A SERVICE OF

logo

3 - 4
No. 99MBC034B
3.1.5 List of parameters
* Factory-set default
No.
Parameter name Setup value: function
00 Reference/Modification 0
*
: For reference only
1: Can be modified
2: Not available
3: Not available
10 Parameter clear
1: Resets to factory-set default
11
Selection of counting direction
If the spindle is pushed-in
(to be set for each INPUT number)
0
*
: + counting
1: - counting
12 Case of EV-16P or EV-16Z 0: 10µm gage
Specification of gage resolution
*1
1: 5µm gage
(to be set for each INPUT number)
2
*
: 1µm gage
3: 0.5µm gage
4: 0.1µm gage
12 Case of EV-16D 0: INC (LGS series)
Specification of gage type 1
*
: ABS (LGD series)
2
*
: ABS ORG
13 Designation of available 6
*
: 6 gages are used
axes
*2
5: 5 gages are used
4: 4 gages are used
3: 3 gages are used
2: 2 gages are used
1: 1 gage is used
14 Start-up mode 0
*
: Counting stand-by
1: Counting execution
15 Display unit
*1
0
*
: mm
*3
1: E (=1/25.4 mm)
17 Designation of calculation 0
*
: CEL1, 2
axis
*4
1: CEL1, 2, 3
2: CEL1, 2, 3, 4
3: All CELs designated as
available axis
*5
20 I/O output mode
*6
0: Command
1
*
: Interval
21
I/O output logic
*7
Tolerance judgment and segment
0
*
: Output terminal is L
1: Output terminal is H
BCD
0
*
: H at output 1
1: L at output 1
22 Selection of I/O function
*6
0
*
: NORMAL mode
1: Calculation mode
2: High-speed mode
23 Selection of I/O type
*6
0
*
: Tolerance judgment
1: Segment output
2: BCD output