7 - 16
No. 99MBC034B
3) Interval mode (for outputting all CELs)
According to the internal timing of the counter, the data of
all CELs (specified by SET1 through SET3) will be output
sequentially.
START
READY
Tolerance judgment, segment, BCD data
CEL1 CEL2 CEL3 CEL4 CEL5 CEL6
30ms - 35ms
5ms - 6ms
1.6ms - 2ms
4) Interval mode (for outputting individual CELs)
According to the internal timing of the counter, the data of
each CEL (specified by SET1 through SET3) will be output
sequentially.
READY
Tolerance judgment, segment, BCD data
30ms - 35ms (5ms)
12ms - 15ms (1.6ms - 2ms)
( ): If outputted in the
high-speed mode
7.7.4 External presetting
Use the current value of a CEL specified by SET1 through
SET3 as a preset value.
SET 1, 2, 3
P. SET
DATA
Min. 10ms Min. 10ms
Max. 40ms
Min. 15ms
If presetting has been executed, the peak values will be cleared.
(Max.= Min. = current value, TIR = 0)