4 - 9
4 BASIC OPERATION
No. 99MBC034B
4.7 Presetting (with the D-EV)
Replace the current value with a preset value.
(1) Press the DISP key to select a CEL number. The count
value of the selected CEL will be displayed (on the D-EV)
or output.
(2) Press the MODE key to select the current-value mode.
The on/off pattern of the LEDs show the selected mode.
[MODE LED Indicators]
MAX MIN
TIR
(3) Press the P.SET key to replace the current value (displayed
value) with a preset value. At this time the peak values will
be cleared and Max. = Min. = Current value, and TIR = 0.
NOTE The EV Counter can count a maximum of eight digits,
however the D-EV Counter can display a maximum of
six digits. If a number has more than six digits, the
most significant one or two digits will be represented
by an "F". If the count value returns to a number of six
digits or less, the D-EV also returns to normal display.
Example)
Count value of EV Counter main unit: 1000.001
Display on the D-EV: F00.001
If the EV-16D is connected to an ABS_ORG type
gage, approximately four seconds is required for the
preset value to be recalled on the EV-16D. During this
time do not move the spindle. Doing so may displace
the zero point.
An ABS or ABS_ORG type gage can be preset one
million times (nominal).