7 - 12
No. 99MBC034B
7.6 High-speed Mode
7.6.1 Overview of the high-speed mode
1) Quickly outputs the CEL data specified by the output CEL
designation.
2) Data output interval in the high-speed mode
5ms (30ms in NORMAL mode)
3) Designating CELs to be output
For the detailed procedure, refer to section 7.7.5 Designating
CELs to be output/Specifying the calculation method.
SET3 SET2 SET1
Output CEL
000 Outputs the data of all CELs in a time-
division output format, as in the
NORMAL mode (power-on default).
001 High-speed output of CEL1 data
*
010 High-speed output of CEL2 data
*
011 High-speed output of CEL3 data
*
100 High-speed output of CEL4 data
*
101 High-speed output of CEL5 data
*
110 High-speed output of CEL6 data
*
111 Specification prohibited.
0="H" 1="L"
* Other CEL output functions are disabled.
NOTE About operation during suspension
• Presetting, peak clear, or zero-setting operation (with
D-EV, I/O, RS-232C) for CELs other than those which
have been specified by SET1 through SET3 will be
suspended, and they become effective only after the
CEL has been specified.
•
Even if the tolerance judgment bar display is active, only
the bars of CELs that have been specified will operate.
•
RS-232C output from CELs other than those which have
been specified by SET1 through SET3 will be the previ-
ous data (existed when the suspension state began).
• If the CELs, excluding that which is displayed on the
D-EV, are specified for output CEL designation, the
display values will be held (locked).
•
If key operation is attempted on the D-EV, the I/O func-
tion will output all CEL data in a time-division format.