4 - 10
No. 99MBC034B
4.8 Tolerance Judgment (with the D-EV)
Three tolerance judgment result stages can be displayed on the
GAGE No. LEDs and output through the I/O connector. (For
more information about the I/O output function refer to chapter
7 I/O FUNCTION.)
(1) Press the DISP key to select a CEL number. The count
value of the selected CEL will be displayed (on the D-EV)
or output.
(2) Press the MODE key to select one of the peak modes: Max.,
Min., and TIR. The GAGE No. LED's on pattern, indicating
the selected CEL number will light in a different color ac-
cording to the tolerance judgment result.
(Display example of tolerance judgment result from CEL1)
+NG:
GAGE
Red
(upper tolerance limit < design value)
GO:
GAGE
Green
(lower tolerance limit design value upper
tolerance limit)
-NG:
GAGE
Amber
(design value < lower tolerance limit)
NOTE There will always be six I/O output channels.
The tolerance judgment result can be also verified
with the 6-axis tolerance judgment bar display. (Refer
to section 4.2 Switching the Display/Output CH (with
the D-EV).