3 - 5
3 SETTING PARAMETERS
No. 99MBC034B
* Factory-set default
No.
Parameter name Setup value: function
25 Baud rate
*8, *9
0: 4800 bps
1
*
: 9600 bps
2: 19200 bps
26 Parity
*8, *9
0: None
1: Odd number
2: Even number
27 Data bits
*8, *9
0
*
: 7 bits
1: 8 bits
28 HOLD selection
*9
0
*
: HOLD
1: RS-232C output
*10
41
Origin detect direction (EV-16Z only)
0*: origin detection at + count
1: origin detection at - count
42
Origin re-detection
*11
(EV-16Z only)
0*: not detect
1: detect
43
Origin initialize (EV-16Z only) Starts with 0 from the origin.
1: initialize (one-shot)
*1
: If this parameter is modified, each zero point, preset value,
and tolerance value for CEL1 to CEL6 will be cleared.
*2
:
For information about this function refer to TIP on page 3-3.
*3
: This is not modified even if an attempt is made to clear the
parameters. If the display unit is set to E, the UNIT LED
lights in green.
*4
: Sets the CEL to be used for the calculation function. (This
function is valid only if Selection of I/O function is set to
Calculation mode.)
*5
: If all CELs designated as the available axis are se-
lected, CELs to be used for calculation will be as shown in
the table below.
Setup value for designating CEL to be used for calculation
the available axes
6 1, 2, 3, 4, 5, 6
5 1, 2, 3, 4, 5
4 1, 2, 3, 4
3 1, 2, 3
2 or 1 1, 2
*6
: For information about this function refer to chapter 7 I/O
FUNCTION.
*7
: The output logic of the data section will vary depending on
the I/O type selection result. However, the logic of the in-
put signal, NORMAL, READY, START, and EXTEND
output remains fixed.
*8
: For information about this function refer to chapter 5 RS-
232C COMMUNICATION FUNCTION.