A SERVICE OF

logo

7 - 5
No. 99MBC034B
7 I/O FUNCTION
No. Name I/O
Functional description
32 OUTCEL IN Sets the output CEL/calculation
33 SET1 IN method specified by SET1 through SET3
34 SET2 IN if the ON signal of OUTCEL was input.
35 SET3 IN In NORMAL/High-speed mode:
CEL designation
In calculation mode:
Specification of calculation method
36 HOLD IN The display value is held during the input.
The error is cleared at the rise of this signal.
Only EV-16Z re-detects the origin if the pa-
rameter No.42 is set to 1.
*: Outputs of No.3 through No.27 can be logically inverted
by the I/O output logic parameter.
**:Inputs of No.31 through No.36 are valid at "L".
4) Designating CELs to be output
For the detailed procedure, refer to section 7.7.5 Designat-
ing CELs to be output/Specifying the calculation
method.
SET3 SET2 SET1
Operation
000Outputs the data of all CELs as a
time-division BCD (power-on de-
fault)
001BCD output of CEL1 data
010BCD output of CEL2 data
011BCD output of CEL3 data
100BCD output of CEL4 data
101BCD output of CEL5 data
110BCD output of CEL6 data
111Specification prohibited.
0="H" 1="L"