7.4.3 BCD output
1) Function
Outputs the CEL data specified by the output CEL designa-
tion as a BCD (6 digits).
2) Parameter setup
I/O function selection parameter:Set to NORMAL
I/O-type selection parameter: Set to BCD output
3) Pin assignment
1
19
18
36
No. Name I/O
Functional description
1
COM
Common terminal of I/O circuit
2 COM
(connected to internal GND)
31×10
0
OUT BCD output data
42×10
0
OUT
54×10
0
OUT
68×10
0
OUT
71×10
1
OUT
82×10
1
OUT
94×10
1
OUT
10
8×10
1
OUT
11
1×10
2
OUT
12
2×10
2
OUT
13
4×10
2
OUT
14
8×10
2
OUT
15
1×10
3
OUT
16
2×10
3
OUT
17
4×10
3
OUT
18
8×10
3
OUT
19
1×10
4
OUT
20
2×10
4
OUT
21
4×10
4
OUT
22
8×10
4
OUT
23
1×10
5
OUT
24
2×10
5
OUT
25
4×10
5
OUT
26
8×10
5
OUT
27
SIGN OUT
Sign of count value
("H" for "+" and "L" for "-")
28
READY
OUT
"L" if the output data has been accepted.
29
START
OUT
"L" if only CEL1 data has been output.
30 NORMAL
OUT
"H" at an anomaly
31
P.SET
IN Presets the specified CELs.