7 - 2
No. 99MBC034B
7.4 NORMAL Mode
7.4.1 Overview of the NORMAL mode
• The NORMAL mode is the factory-set default. Use it with-
out modification for general purposes.
• Origin setting, tolerance setting, and presetting can be per-
formed individually for each CEL.
• Tolerance judgment (for each CEL), BCD data, and seg-
ment (on selected CELs) output can be performed.
7.4.2 Tolerance judgment output
1) Function
Outputs the tolerance judgment result of each internal
counter (CEL1 to CEL6).
Measured value < Lower tolerance limit -NG output
Lower tolerance limit ≤ Measured value ≤ Upper tolerance limit
GO output
Upper tolerance limit < Measured value +NG output
2) Parameter setup
I/O function selection parameter:Set to NORMAL
I/O-type selection parameter: Set to tolerance judgment
3) Pin assignment
1
19
18
36
No. Name I/O
Functional description
1
COM Common terminal of I/O circuit
2
COM (connected to internal GND)
3
CEL1_-NG OUT
Tolerance judgment result of CEL1
-NG
4
CEL1_GO OUT
Tolerance judgment result of CEL1
GO
5
CEL1_+NG OUT
Tolerance judgment result of CEL1
+NG
6 CEL1_NOM OUT
Error signal of CEL1
*1
7 CEL2_-NG OUT
Tolerance judgment result of CEL2
-NG
8 CEL2_GO OUT
Tolerance judgment result of CEL2
GO
9
CEL2_+NG OUT
Tolerance judgment result of CEL2
+NG
10 CEL2_NOM OUT
Error signal of CEL2
*1
11 CEL3_-NG OUT
Tolerance judgment result of CEL3
-NG
12 CEL3_GO OUT
Tolerance judgment result of CEL3
GO
13 CEL3_+NG OUT
Tolerance judgment result of CEL3
+NG
14 CEL3_NOM OUT
Error signal of CEL3
*1
15 CEL4_-NG OUT
Tolerance judgment result of CEL4
-NG
16 CEL4_GO OUT
Tolerance judgment result of CEL4
GO
17 CEL4_+NG OUT
Tolerance judgment result of CEL4
+NG
18 CEL4_NOM OUT
Error signal of CEL4
*1