5 - 5
5 RS-232C COMMUNICATION FUNCTION
No. 99MBC034B
5.2.5 RS-232C command and RS-232C response output
LF
Command
Response
output
Max. 10ms
(
If the GA command is issued
)
Min. 10ms
LF
LF
NOTE During key operation RS-232C output will be sus-
pended, and it will be resumed after the key operation
has been completed.
5.2.6 HOLD input and RS-232C response output
LF
HOLD
EXTEND
Max. 70ms
Min. 10ms
Min. 10ms
Data latch for
10 µs [10 ms]*
* A value in brackets denotes a numeric value
at a MAX, MIN, or TIR measurement.
Response
output
NOTE During key operation RS-232C output will be sus-
pended, and it will be resumed after the key operation
has been completed.
If the RS LINK connection has been established, the
EXTEND command is valid for the counter placed at
the end of the chain.
EXTEND output will be in effect only if the I/O mode
parameter is set to "Interval mode".
While the HOLD signal is being input, the D-EV UNIT
LED flashes.
In order to trigger data output from the RS-232C con-
nector with the HOLD input signal set the HOLD se-
lection parameter to "RS-232C". (In this state an RS-
232C command cannot be input.)
If the High-speed mode is specified, set the output cell
to "All CELs".