3 - 6
No. 99MBC034B
*9
: The parameter modification will be made valid after reset-
ting the power.
*
10
: This function will output data from the RS-232C connector
if a HOLD signal is input through the I/O connector. In this
case, all RS-232C commands will be ineffective.
*
11
: For details, refer to the section, "2.3 Power On".
3.2 Setting Parameters (with EV Counter)
Before using the EV Counter, various parameters must be set
up. The parameter setup operation is easily performed if a D-
EV is used.
IMPORTANT The EV-16Z can not modify the origin-related set-
tings by itself. For making this modification in EV-
16Z, prepare a D-EV.
NOTE • Use a D-EV for the resolution of 0.1µm setting.
Set up of EV Counter alone, is not possible.
• Only a D-EV allows setting when the modification is
made for the parameter no.41, 42, and 43.
• In order to connect a gage with a resolution that is not
1µm, the counter must modify its resolution setting to
that of the connected gage if the EV-16P or EV-16Z is
used.
• The gage type (INC, ABS, etc.) must be set for the
EV-16D.
• If the setup is made as (Number of available axes) >
(Number of gages), an error display/output (Error40)
will result.
If the setup is made as (Number of available axes) <
(Number of gages), gages that are not designated will
be ignored. For information refer to section 8.4 List of
Error Outputs.
3.2.1 Parameter mode ON
Set the parameter mode to ON to modify the existing parameter
settings.
(1) Press the LOAD button while holding down the DATA but-
ton to set the parameter mode to ON.
[Parameter LED Indicator]
87 6
54
321
LOAD MODE SEL. DATA
(2) Begin setting the parameters.