7 I/O FUNCTION
7.7.3 Data output
There are two data output methods that can be specified by the
I/O output mode parameter: the command mode and the inter-
val mode.
1) Command mode (for outputting all CELs)
With synchronized control of the HOLD and READY sig-
nals, the data of all CELs (specified by SET1 through SET3)
will be output.
HOLD
START
READY
Tolerance judgment, segment, BCD data
Data latch for 10 ยตs (10 ms at a peak measurement), for EV-16P and EV-16Z.
Min. 70ms Min. 15ms
Max. 40ms
5ms - 6ms
1.6ms - 2ms
CEL1 CEL2 CEL3 CEL4 CEL5 CEL6
*While the HOLD signal is being input, the UNIT LED (on D-
EV) will flash.
2) Command mode (for outputting individual CELs)
With synchronized control of the HOLD and READY sig-
nals, the data of each CEL (specified by SET1 through
SET3) will be output.
HOLD
READY
Tolerance judgment, segment, BCD data
Data latch for 10 _s (10 ms at a peak measurement), for EV-16P and EV-16Z.
Min. 40ms (Min. 15ms) [Min. 70ms]
Max. 40ms
(Max. 15ms)
[Max. 80ms]
Min. 15ms
12ms - 15ms
(1.6ms - 2ms)
( ): If outputted in the
high-speed mode.
[ ]: If outputted in the
calculation mode.
NOTE If this command is used in the high-speed mode or in the
all-CELs output mode, use a device that has an input
response time that is less than 1ms.