A SERVICE OF

logo

4 - 8
No. 99MBC034B
4.6 Clearing the Error State
The error state can be cleared with the following procedures.
If the EV-16Z is used, push in the spindle to pass through the
origin after completing the following procedures.
(1) Press the DISP key to select a CEL number in the error
state. Then press P.SET key.
(2) The HOLD input signal rises to "H". (I/O)
(3) Input the CS command. (RS-232C)
NOTE While the error state is being canceled, Error40 is dis-
played (on D-EV) or output.
While the error state is being canceled, the GAGE No.
LED flashes.
Time required for error cancelation: 30ms (EV-16P
and EV-16Z), max. 8s ( EV-16D)