7 - 10
No. 99MBC034B
7.5 Calculation Mode
7.5.1 Overview of the calculation mode
1) Using the CEL that has been specified by the calculation
axis designation parameter, output the calculation result
specified by the input of SET1 to SET3, as shown below.
2) Setting the calculation method
(For the detailed procedure, refer to section 7.7.5 Designating
CELs to be output/Specifying the calculation method .)
SET3 SET2 SET1
Calculation method
000 No calculation (power-on default)
001
Sum of CEL data on specified calculation axes
010
Mean of CEL data on specified calculation axes
011
Max. of CEL data on specified calculation axes
100
Min. of CEL data on specified calculation axes
101
MAX - MIN of CEL data on specified calculation
axes
110 Specification prohibited.
111 Specification prohibited.
0="H" 1="L"
7.5.2 Tolerance judgment result output
1) Function
• A tolerance judgment of the calculation result will be out-
put from CEL6.
• In case of no calculation, a tolerance judgment to the
count value of CEL6 will be output.
• The tolerance limits for CEL6 will be applied.
2) Parameter setup
•
I/O function selection parameter: Set to the calculation mode.
• I/O-type selection parameter: Set to tolerance judgment.
• Calculation axis designation parameter:
Select one between 0 and 3.
TIP • Display on the D-EV when the calculation mode is be-
ing set.
If the display is switched with the DISP key, the follow-
ing will be sequentially displayed after normal display
(CH01 to CH06 and 6-axis tolerance judgment bar):
Displays the sum after displaying CL01_1.
Displays the mean after displaying CL02_2.
Displays the maximum value after displaying CL03_3.
Displays the minimum value after displaying CL04_4.
Displays the MAX - MIN after displaying CL05_5.
For the second EV Counter on the RS LINK, the fol-
lowing will be displayed: