Agilent Technologies 86120B Marine Instruments User Manual


 
1-2
Getting Started
Getting Started
Getting Started
The instructions in this chapter show you how to install your Agilent 86120B.
You should be able to finish these procedures in about ten to twenty minutes.
After you’ve completed this chapter, continue with Chapter 2, “Using the
Multi-Wavelength Meter”. Refer to Chapter 7, “Specifications and Regulatory
Information” for information on operating conditions such as temperature.
If you should ever need to clean the cabinet, use a damp cloth only.
WARNING
To prevent electrical shock, disconnect the Agilent 86120B from
mains before cleaning. Use a dry cloth or one slightly dampened with
water to clean the external case parts. Do not attempt to clean
internally.
CAUTION
Ventilation Requirements. When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4
°
C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
then forced convection must be used.
CAUTION
This product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.
CAUTION
Install the instrument so that the ON/OFF switch is readily identifiable and is
easily reached by the operator. The ON/OFF switch or the detachable power
cord is the instrument disconnecting device. It disconnects the mains circuits
from the mains supply before other parts of the instrument. Alternately, an
externally installed switch or circuit breaker (which is really identifiable and is
easily reached by the operator) may be used as a disconnecting device.
CAUTION
Install the instrument according to the enclosure protection provided. This
instrument does not protect against the ingress of water. This instrument
protects against finger access to hazardous parts within the enclosure.