Agilent Technologies 86120B Marine Instruments User Manual


 
5-42
Programming Commands
CALCulate2 Subsystem
WLIMit:STOP:WNUMber
Sets the stopping wavenumber for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP:WNUMber {?|{<real> | MINimum | MAXimum }}
<
real
> is a wavenumber value that is within the following limits:
Constant
Description
MINimum start wavelength limit
MAXimum 14286 cm
-1
(700 nm)
Attribute
Summary
Non-sequential command
Preset
State: 8.333335E5 m
-1
*RST State: 8.333335E5 m
-1
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The default
units for the <
real
> parameter are inverse meters. The stop wavenumber
value must be less than or equal to the start wavenumber value or the stop
wavenumber will be clipped to the start wavenumber and a “Data out of
range” error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument. Refer to
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.