5-85
Programming Commands
TRIGger Subsystem
INITiate:CONTinuous
Selects single or continuous measurement acquisition.
Syntax
:INITiate:CONTinuous{?| {ON | OFF | 1 | 0}}
Attribute
Summary
Non-sequential command
Preset
State: on
*RST State: off
SCPI Compliance: standard
Description
When on is specified, the instrument continuously measures the input spec-
trum.
INITiate[:IMMediate]
Initiates a new measurement sequence.
Syntax
:INITiate:IMMediate
Attribute
Summary
Non-sequential command
Preset
State: none
SCPI Compliance: standard
Command Only
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument. Refer to
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.