Agilent Technologies 86120B Marine Instruments User Manual


 
5-84
Programming Commands
TRIGger Subsystem
TRIGger Subsystem
The SCPI definition defines the TRIGger subsystem to include ABORt, ARM,
INITiate, and TRIGger commands. The Agilent 86120B has no ARM or TRIG-
ger commands.
The commands in this subsystem have the following command hierarchy:
ABORt
INITiate
:CONTinuous
[:IMMediate]
ABORt
Halts the current measurement sequence and places the instrument in the
idle state.
Syntax
:ABORt
Attribute
Summary
Preset
State: not affected
SCPI Compliance: standard
Command Only
Description
If the instrument is configured for continuous measurements, a new measure-
ment sequence will begin. Otherwise, the instrument stays in the idle state
until a new measurement is initiated.