Agilent Technologies 86120B Marine Instruments User Manual


 
5-39
Programming Commands
CALCulate2 Subsystem
WLIMit:STARt:WNUMber
Sets the starting wavenumber for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STARt:WNUMber {?|{<real> | MINimum | MAXimum}}
<
real
> is a wavenumber value that is within the following limits:
Constant
Description
MINimum 6060 cm
-1
MAXimum wavelength limit stop value
Attribute
Summary
Non-sequential command
Preset
State: 6.060606E5 m
-1
*RST State: 6.060606E5 m
-1
SCPI Compliance: instrument specific
Description
This command sets the starting range for the wavelength limit. The default
units for the <
real
> parameter value are m
-1
. The start wavenumber value
must be less than or equal to the stop wavenumber value or the start wave-
number will be clipped to the stop wavenumber and a “Data out of range”
error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument. Refer to
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.