Agilent Technologies 86120B Marine Instruments User Manual


 
5-40
Programming Commands
CALCulate2 Subsystem
WLIMit:STOP:FREQuency
Sets the stopping frequency for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP:FREQuency {?|{<real> | MINimum | MAXimum }}
<
real
> is a frequency value that is within the following limits:
Constant
Description
MINimum start wavelength limit
MAXimum 428.2750 THz
Attribute
Summary
Non-sequential command
Preset
State: 249.8271 THz
*RST State: 249.8271 THz
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The default
units for the <
real
> parameter are Hz. The stop frequency value must be
greater than or equal to the start frequency value or the stop frequency will be
clipped to the start frequency and a “Data out of range” error will be gener-
ated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument. Refer to
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.