CORRection:ELEVation
Sets the elevation value used by the instrument to compensate for air disper-
sion.
Syntax
:SENSe:CORRection:ELEVation{?| {<integer> | MINimum | MAXimum}}
<integer>
is the altitude in meters.
Constant
Description
MINimum 0 m
MAXimum 5000 m
Attribute
Summary
Non-sequential command
Preset
State: unaffected by
*RST sets this value to the minimum.
SCPI Compliance: instrument specific
Description
Changing the elevation value causes the current data to be reprocessed.
Query Response
The query form returns the current elevation setting as shown in the following
example:
+1500
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument. Refer to
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.