Step 2. Check the Fuse
Step 2. Check the Fuse
1
Locate the line-input connector on the instrument’s rear panel.
2
Disconnect the line-power cable if it is connected.
3
Use a small flat-blade screwdriver to open the pull-out fuse drawer.
4
Verify that the value of the line-voltage fuse in the pull-out drawer is correct.
The recommended fuse is an IEC 127 5
×
20 mm, 6.3A, 250 V, Agilent
Technologies part number 2110-0703.
Notice that an extra fuse is provided in a drawer located on the fuse holder.
WARNING
For continued protection against fire hazard, replace line fuse only
with same type and ratings, (type T 6.3A/250V for 100/240V
operation). The use of other fuses or materials is prohibited.