Agilent Technologies 86120B Marine Instruments User Manual


 
5-74
Programming Commands
STATus Subsystem
Query Response
The following string shows an example of the first few measurements returned
by this query.
+1.51367200E+000,+1.51855500E+000,+1.49902300E+000,+1.47949200E+000,+1.50488300E+000,+1.5
3320300E+000,+1.50097700E+000,+1.47265600E+000,+1.50293000E+000,+1.50781300E+000,+1.51171
900E+000,+1.48242200E+000,+1.50097700E+000,+1.51855500E+000,+1.50683600E+000,+1.48632800E
+000,+1.50488300E+000
Notice that only values are returned to the computer. There is no first value
that indicates the number of values contained in the string as there is, for
example, with the FETCh, READ, and MEASure commands.
STATus Subsystem
Use the commands in this subsystem to control the Agilent 86120B’s status-
reporting structures. These structures provide registers that you can use to
determine if certain events have occurred.
The commands in this subsystem have the following command hierarchy:
:STATus
:OPERation
:CONDition?
:ENABle
[:EVENt]?
:PTRansition
:NTRansition
:PRESet
:QUEStionable
:CONDition?
:ENABle
[:EVENt]?
:PTRansition
:NTRansition