Agilent Technologies 86120B Marine Instruments User Manual


 
5-54
Programming Commands
CALCulate3 Subsystem
DRIFt:MAXimum[:STATe]
Sets the drift calculation to return the maximum power and frequency values
measured.
Syntax
:CALCulate3:DRIFt:MAXimum[:STATe]{?| {ON | OFF | 1 | 0}}
Attribute
Summary
Preset
State: off
*RST State: off
SCPI Compliance: instrument specific
Description
Use the CALC3:DRIF:PRES command to turn off all the drift states before
turning on this state. The CALC3:DATA? query returns the maximum power
and frequency.
Note
Only one STATe command can be turned on at any one time. Attempting to turn more
than one state on at a time results in a “–221 Settings Conflict” error. Refer to “Measure
delta, drift, and signal-to-noise” on page 4-14 for additional information on selecting
measurements.