Agilent Technologies 86120B Marine Instruments User Manual


 
6-4
Performance Tests
Test 2. Sensitivity
Test 2. Sensitivity
Description
Sensitivity is verified using the following devices:
Optical power meter
Optical attenuator
1310 nm and 1550 nm lasers (>0 dBm output power)
CAUTION
Do not exceed +18 dBm source power. The Agilent 86120B’s input circuitry can
be damaged when total input power exceeds 18 dBm.
Procedure
Perform the following procedure first using the 1310 nm laser and then repeat
the steps using the 1550 nm laser.
1
Connect the laser’s output to the optical attenuator’s input.
2
Connect the optical attenuator’s output to the optical power meter.
3
Adjust the attenuator for a reading of 0 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at 0 dBm: ____________
4
Adjust the attenuator for a reading of –35 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at –35 dBm: ____________
5
Disconnect the fiber-optic cable at the power meter’s input and connect the
cable to the Agilent 86120B being tested.
6
Reset the optical attenuator to the setting recorded in Step 3.
7
Read the power and wavelength measured on the Agilent 86120B, and
compared them to the specifications listed in Chapter 7, “Specifications and
Regulatory Information”.
8
Reset the optical attenuator to the setting recorded in Step 4.
9
Read the power and wavelength measured on the Agilent 86120B, and
compared them to the specifications listed in Chapter 7, “Specifications and
Regulatory Information”.