Locking or unlocking the meter
After you initialize the meter, you must lock it in order to conform to MID standards.
1. Navigate to Maint > Setup > Meter >Lock.
2. Press Edit to activate or deactivate the lock.
3. Enter your lock password.
NOTE: The default password is 0000. To set up a new password see “Setting up
lock password” on page 112.
4. Press + and - to scroll between Active and Inactive.
5. Press OK to select the option.
6. Select Yes to confirm the selected option, and exit the screen.
On activating the lock, a lock icon appears on the upper left corner of the screen.
7. Make sure you record and store the lock password in a secure location. A lost lock
password cannot be recovered.
Setting up lock password
To set up a new lock password, perform the below procedure. To change the
password, make sure the lock is inactive and perform the below procedure.
NOTE: You cannot change the lock password when the lock is active.
1. Navigate to Maint > Setup > HMI > Pass.
2. Press to scroll to Revenue Lock in the Passwords screen.
3. Press Edit to select a password.
4. Press + to increment the active digit through the numerals 0-9.
5. Press to move to the next digit to the left.
6. Continue until all values are selected, and then press OK to set the password.
7. Press Yes to save the changes.
NOTICE
IRRECOVERABLE PASSWORD
Record your meter's lock password information in a secure location.
Failure to follow this instruction can result in a permanently locked meter.
>
>