Section 5. Service and Troubleshooting
5.6.2 Recommended
Actions
The FD_RECOMMEN_ACT and RECOMMENDED_ACTION parameters display text strings that
will give a recommended course of action to take based on which type and which specific event of
the alert that is active. Table 5-11 provides recommended actions for Field Diagnostic Alerts as
given by the alert default setting for the Rosemount 2230 Display.
Table 5-11. Recommended
actions for Field Diagnostics
Alerts.
Alert Type Description Recommended Action
Failure
Software Failure 1. Restart the device.
2. Load default database to the
device and reconfigure the device.
3. Contact Rosemount Tank Gauging
Service Department.
Database Error 1. Restart the device.
2. Load default database to the
device.
3. Reconfigure the device.
Memory Failure - FF I/O Board 1. Perform a Factory Reset - FF I/O
Board.
2. If error persists, it may indicate a
faulty memory chip. Replace the
transmitter head.
Electronics failure - Main Board Replace the device.
Internal Communication Failure 1. Restart the device
2. Replace the device.
Electronics failure - FF I/O Board Replace the device.
Out of Specification
MAO Fault State Mode Enabled.
One or several MAO blocks are
configured with fault-state mode enabled.
Turn fault state mode off in MAO
block
Internal Temperature Out of Limits Check ambient temperature at
installation site
Invalid Model Code Contact Rosemount Tank Gauging
Service Department.
Function Check Check Function.
One or more Transducer Blocks are in
Out of Service mode.
Return transducer block to Auto mode