Emerson 2230 Marine Sanitation System User Manual


 
Reference Manual
00809-0100-2230, Rev BB
August 2014
Rosemount 2230
A-2
Appendix A. Reference Data
FOUNDATION™ fieldbus characteristics
Polarity sensitive No
Quiescent current draw 30 mA
Lift-off minimum voltage 9.0 VDC
Device capacitance / inductance See Appendix B: Product Certifications
Class (Basic or Link Master) Link Master (LAS)
Number of available VCRs Maximum 38. Client and server=20, Publisher=20, Subscribers=20, Source=2, Sink=0.
Links Maximum 32
Minimum slot time / maximum response delay
/ minimum intermessage delay
8 /5 /8
Blocks and Execution time 1 Resource block, 3 Transducer blocks (Main, Register, Display)
4 Multiple Analog Output (MAO) blocks: 15 ms,
For more information, see the FOUNDATION™ fieldbus Blocks manual (document number
00809-0100-4783)
Instantiation No
Conforming FOUNDATION™ fieldbus ITK 6
Field Diagnostics support (NAMUR 107) Yes
Action support wizards Write protect device, factory reset - device configuration, reset statistics, start/stop alerts
simulation, restart communication
Advanced diagnostics Software, memory/database, electronics, internal communication, configuration, model
code, internal temperature, MAO fault state
Mechanical
Housing material Polyurethane-covered die-cast aluminum
Dimensions (width x height x depth) 150 x 120 x 78 mm (5.9 x 4.7 x 3.1 in.)
Weight 1.2 kg (2.6 lbs)
Environment
Ambient temperature -20 to 70 °C (-4 to 158 °F)
Storage temperature -30 to 85 °C (-22 to 185 °F)
Humidity 0-100% relative humidity, non-condensing
Ingress protection IP 66 and 67 (Nema 4)
Metrology sealing possibility Yes
Write protect switch Yes
(1) Make sure that unused ports are properly sealed to prevent moisture or other contamination from entering the electronics housing. Use the enclosed metal
plug for this purpose.