HP (Hewlett-Packard) E1412A Waterskis User Manual


 
Multimeter Command Reference 101Chapter 3
MEASure
The MEASure command subsystem configures the multimeter to perform the
specified measurement with the given range and resolution. When the multimeter is
triggered,
MEASure makes the measurement and sends the readings to the output
buffer.
Executing
MEASure is equivalent to configuring the multimeter with the low-level
commands shown in the following table.:
Subsystem Syntax MEASure
:CURRent:AC? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:CURRent[:DC]? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:FREQuency? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:FRESistance? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:PERiod? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:RESistance? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:VOLTage:AC? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
[:VOLTage[:DC]]? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
[:VOLTage[:DC]]:RATio? [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
Command Setting
RANGe As specified (or AUTO).
RESolution As specified, or as a function of range, integration
time, or NPLCs.
AC filter
([SENSe:]DET:BAND)
20 Hz - 300 kHz (medium filter)
Autozero
([SENSe:]ZERO:AUTO)
OFF if resolution setting results in NPLC <1;
ON if resolution setting results in NPLC 1
Input resistance
([SENSe:]INP:IMP:AUTO)
Applies to dc voltage and is disabled for all other
functions. 10M for all dc voltage ranges.
Samples per trigger
(SAMP:COUN)
Trigger count
(TRIG:COUN)
Trigger delay
(TRIG:DEL)
Trigger source
(TRIG:SOUR)
1 sample
1 trigger
AUTO (Automatic delay)
IMM (trigger signal is always true)
Math function
(CALCulate:STATe)
OFF