Multimeter Command Reference 85Chapter 3
CONFigure
The CONFigure command subsystem configures the multimeter to perform the
specified measurement with the given range and resolution.
CONFigure does not
make the measurement after setting the configuration. Executing
CONFigure is
equivalent to setting the multimeter configuration as follows:
After configuring the multimeter, use the
INITiate command to place the multimeter
in the wait-for-trigger state and store readings in the multimeter’s internal memory.
Or, use the
READ? command to make the measurement and send the readings to the
output buffer when the trigger is received.
Subsystem Syntax CONFigure
:CURRent:AC [<range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:CURRent[:DC] [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:FREQuency [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:FRESistance [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:PERiod [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:RESistance [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
:VOLTage:AC [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
[:VOLTage[:DC]] [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
[:VOLTage[:DC]]:RATio [<
range>|MIN|MAX|DEF|AUTO[,<resolution>|MIN|MAX|DEF]]
The CONFigure command RANGe and RESolution parameters are optional. You
will get the default range and resolution settings if you do not specify a range or
resolution in the command. You will get these default settings even if you set a range
Command Setting
RANGe As specified (or AUTO).
RESolution As specified, or as a function of range, integration
time, or NPLCs.
AC filter
([SENSe:]DET:BAND)
20 Hz - 300 kHz (medium filter)
Autozero
([SENSe:]ZERO:AUTO)
OFF if resolution setting results in NPLC <1;
ON if resolution setting results in NPLC 1
Input resistance
([SENSe:]INP:IMP:AUTO)
Applies to dc voltage and is disabled for all other
functions. 10M for all dc voltage ranges.
Samples per trigger
(SAMP:COUN)
Trigger count
(TRIG:COUN)
Trigger delay
(TRIG:DEL)
Trigger source
(TRIG:SOUR)
VM Complete routing
(OUTP:TTLT<
n>:STAT)
1 sample
1 trigger
AUTO (Automatic delay)
IMM (trigger signal is always true)
OFF (all trigger lines;
n = 0 - 7)
Math function
(CALCulate:STATe)
OFF