190 HP E1312A and HP E1412A Multimeter Error Messages Appendix B
621 AC rms full scale failed
622 Frequency counter failed
623 Cannot calibrate precharge
625 I/O processor does not respond
626 I/O processor failed self-test
Calibration Errors The following errors indicate failures that may occur during a calibration.
The most common errors have descriptions here. Refer to the HP E1312A
and HP E1412A Service Manual for more information on the other errors.
701 Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the
multimeter. When applicable, this error will occur at power-on to warn you
that the multimeter is unsecured.
702 Cal secured
The multimeter is secured against calibration.
703 Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the multimeter. You must use the same security code to
unsecure the multimeter as was used to secure it, and vice versa. The
security code may contain up to 12 alphanumeric characters. The first
character must be a letter.
704 Secure code too long
A security code was received which contained more than 12 characters.
705 Cal aborted
A calibration in progress is aborted when you send a device clear to the
multimeter.
706 Cal value out of range
The specified calibration value (
CAL:VALue) is invalid for the present
function and range.
707 Cal signal measurement out of range
The specified calibration value (
CAL:VALue) does not match the signal
applied to the multimeter.
708 Cal signal frequency out of range
The input signal frequency for an ac calibration does not match the required
input frequency for calibration.