188 HP E1312A and HP E1412A Multimeter Error Messages Appendix B
-330 Self-test failed
The multimeter’s complete self-test failed from the remote interface (
*TST?
command). In addition to this error, more specific self-test errors are also
reported. See also “Self-Test Errors,” following this section.
-350 Too many errors
The error queue is full because more than 20 errors have occurred. No
additional errors are stored until you remove errors from the queue. The
error queue is cleared when power has been off, or after a
*CLS (clear status)
command has been executed.
-410 Query INTERRUPTED
A command was received which sends data to the output buffer, but the
output buffer contained data from a previous command (the previous data is
not overwritten). The output buffer is cleared when power has been off, or
after a
*RST (reset) command has been executed.
-420 Query UNTERMINATED
The multimeter was addressed to talk (i.e., to send data over the interface)
but a command has not been received which sends data to the output buffer.
For example, you may have executed a
CONFigure command (which does
not generate data) and then attempted an
ENTER statement to read data from
the remote interface.
-430 Query DEADLOCKED
A command was received which generates too much data to fit in the output
buffer and the input buffer is also full. Command execution continues but all
data is lost.
-440 Query UNTERMINATED after indefinite response
The
*IDN? command must be the last query command within a command
string. Example:
*IDN?;:SYST:VERS?
501 Isolator UART framing error
502 Isolator UART overrun error
511 Unexpected reset occurred
The outguard circuit recognized the inguard circuit reset (probably due to an
abnormal input condition). This error causes the instrument to go to the
power-on setting and the previous setting is lost.
521 Input buffer overflow
522 Output buffer overflow
531 Insufficient memory
There is not enough memory to store the requested number of readings in
internal memory using the
INITiate command. The product of the sample
count (
SAMPle:COUNt) and the trigger count (TRIGger:COUNt) must not
exceed 512 readings.