92 Multimeter Command Reference Chapter 3
:RESistance
CONFigure:RESistance [<range>|MIN|MAX|DEF|AUTO
[,<
resolution>|MIN|MAX|DEF]] selects the 2-wire ohms function and allows you to
specify the range and resolution.
Parameters
Comments To select a standard measurement range, specify range as the input signal’s
maximum expected resistance. The multimeter then selects the correct range
that will accept the input.
The AUTO or DEFault option for the range parameter enables autorange. The
DEFault option for resolution defaults the integration time to 10 PLC.
The MIN and MAX parameters select the minimum or maximum values for
range and resolution:
For range:
MIN = 100
; MAX =100M
For resolution:
MIN selects the best resolution (the smallest value) for the
selected range.
MAX selects the worst resolution (the largest value) for the
selected range.
To select autorange, specify DEF for range or do not specify a value for the
range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
To specify a MIN or MAX resolution while autoranging, you must specify
AUTO or DEFault for range; CONF:RES DEF,MIN or CONF:RES DEF,MAX
or
CONF:RES AUTO,MIN or CONF:RES AUTO,MAX (you cannot omit the
range parameter). This prevents the
MIN or MAX resolution from being
interpreted as a range setting.
Related Commands: FETCh?, INITiate, READ?
Example Making 2-Wire Ohms Measurements
CONF:RES 850,MAX
Function: 2-wire ohms; range selected: 1k
;
MAX resolution: 0.1.
SAMP:COUN 3
Take 3 readings.
INIT
Place multimeter in wait-for-trigger state;
store readings in internal memory; trigger
source is IMMediate by default.
FETC?
Place readings in output buffer.
enter statement
Enter readings into computer.
Parameter Name Parameter Type Range of Values Default Units
<
range>
numeric 100Ω|1k|10k| 100k|1M|
10M|100M|
MIN|MAX|DEF|AUTO
ohms
<
resolution>
numeric
resolution|MIN|MAX|DEF
ohms