Multimeter Command Reference 145Chapter 3
Example Query the Measurement Range
VOLT:AC:RANG 10
Select 10V range.
VOLT:AC:RANG?
Query multimeter to return the present range.
enter statement
Enter value into computer.
VOLTage:AC:RANGe:AUTO
[SENSe:]VOLTage:AC:RANGe:AUTO <mode> enables or disables the autorange
function for AC voltage measurements.
Parameters
Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
If you explicitly select a range using VOLTage:AC:RANGe autoranging is
turned
OFF.
In order to specify an aperture time of 10
µ
s, you must select a fixed range
(e.g.,
VOLT:RANG:AUTO OFF).
Related Commands: CONFigure, VOLTage:RANGe
*RST Condition: VOLT:AC:RANG:AUTO ON
Example Disable AC Voltage Autoranging
VOLT:AC:RANG:AUTO OFF
Disable autorange.
VOLTage:AC:RANGe:AUTO?
[SENSe:]VOLTage:AC:RANGe:AUTO? returns a number to show whether the AC
voltage autorange mode is enabled or disabled: “
1” = ON, “0” = OFF. The value is
sent to the output buffer.
Example Query the Autorange Mode
VOLT:AC:RANG:AUTO OFF
Disable autorange.
VOLT:AC:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<mode>
boolean OFF|0|ON|1 none