Multimeter Command Reference 103Chapter 3
:CURRent[:DC]?
MEASure:CURRent[:DC]? [<range>|MIN|MAX|DEF|AUTO
[,<
resolution>|MIN|MAX|DEF]] selects the DC current function and allows you to
specify the measurement range and resolution.
Parameters
Comments To select a standard measurement range, specify range as the input signal’s
maximum expected current. The multimeter then selects the correct range that
will accept the input.
The AUTO option for the range parameter enables autorange. The DEF option
for the resolution parameter defaults the integration time to 10 PLC.
The DEFault option for the range parameter will also enable autorange.
The
DEF option for the resolution parameter defaults the integration time to
10 PLC.
The MIN and MAX parameters select the minimum or maximum values for
range and resolution:
For range:
MIN = 10mA; MAX = 3A
For resolution:
MIN selects the best resolution (the smallest value) for the
selected range.
MAX selects the worst resolution (the largest value) for the
selected range.
To select autorange, specify DEF for range or do not specify a value for the
range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
To specify a MIN or MAX resolution while autoranging, you must specify the
AUTO or DEFault parameter for range as MEAS:CURR:DC? AUTO or
MEAS:CURR:DC? DEF (you cannot omit the range parameter). This prevents
the
MIN or MAX resolution from being interpreted as a range setting and the
resulting command becomes
MEAS:CURR:DC? DEF,MIN or
MEAS:CURR:DC? DEF,MAX (or use AUTO in place of DEF).
Related Commands: FETCh?, INITiate, READ?
Example Making DC Current Measurements
MEAS:CURR:DC? .1,MAX
Function: DC current; range selected: 1A
MAX resolution: 1.0E-05A.
Parameter Name Parameter Type Range of Values Default Units
<
range>
numeric 10mA|100mA|1 A|3 A|
MIN|MAX|DEF|AUTO
A
<resolution>
numeric
resolution|
MIN|MAX|DEF
A