Multimeter Command Reference 133Chapter 3
FRESistance:RANGe
[SENSe:]FRESistance:RANGe <range> selects the range for 4-wire resistance
measurements.
Parameters
Comments To select a standard measurement range, specify range as the input signal’s
maximum expected resistance. The multimeter then selects the correct range.
MIN selects the minimum range available with the FRESistance:RANGe
command: 100
.
MAX selects the maximum range available: 100
ΜΩ
.
You must select a range using FRESistance:RANGe before specifying
resolution.
Specifying a fixed range disables the autorange mode set by the
FRES:RANG:AUTO command.
The FRES:RANG command overrides the range setting from a previous
CONFigure command on the same function. The multimeter uses the same
aperture time to set the resolution on the new range as was selected by
CONFigure.
*RST Condition: FRES:RANG 1k
(1E+03)
Example Set Four-Wire Resistance Range to 10M
FRES:RANG 1E+07
4-wire resistance range is 10M
.
FRESistance:RANGe?
[SENSe:]FRESistance:RANGe? [MIN|MAX] returns one of the following numbers
to the output buffer:
The present resistance range selected if MIN or MAX is not specified.
The minimum resistance range available (100) if MIN is specified.
The maximum resistance range available (100M) if MAX is specified.
Example Query the Measurement Range
FRES:RANG 100
Select 100 range.
FRES:RANG?
Query multimeter to return the present range.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
range>
numeric 100|1k|10k|100k|1M
|10M|100M|MIN|MAX
ohms