152 Multimeter Command Reference Chapter 3
Example Query the Resolution
VOLT:DC:RES 1E-03
Set resolution to 1 mV.
VOLT:DC:RES?
Query multimeter to return the present resolution.
enter statement
Enter value into computer.
ZERO:AUTO
[SENSe:]ZERO:AUTO <mode> enables or disables the autozero mode. Autozero
applies to dc voltage, dc current and 2-wire ohms measurements only. 4-wire ohms
and dc voltage ratio measurements automatically enable the autozero mode.
Parameters
Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
The ON parameter enables autozero. This is the default parameter which
causes the multimeter to internally disconnect the input signal following each
measurement and make a zero measurement. The zero reading is subtracted
from the input signal reading to prevent offset voltages present on the
multimeter’s input circuitry from affecting measurement accuracy.
The OFF parameter disables autozero. In this mode the multimeter takes one
zero measurement and subtracts it from all subsequent input signal
measurements prior to a change in function, range or integration time. A new
zero measurement is made following a change in function, range or integration
time. This mode increases measurement speed because a zero measurement is
not made for each input signal measurement.
Autozero ONCE issues an immediate zero measurement and can be used to get
an update on the zero measurement for a specific input signal measurement.
This helps to increase measurement time since you update the zero reading
without making zero measurements for every measurement.
*RST Condition: ZERO:AUTO ON (enables autozero mode)
Example Disable Autozero
ZERO:AUTO OFF
Autozero disabled.
ZERO:AUTO?
[SENSe:]ZERO:AUTO? queries the autozero mode. Returns one of the following
responses to the output buffer:
0” (OFF or ONCE) if autozero is disabled or set for one time.
1ON if autozero is enabled.
Example Query the Autozero Mode
ZERO:AUTO?
Queries the autozero mode.
enter statement
Enter response into computer.
Parameter Name Parameter Type Range of Values Default Units
<mode>
boolean OFF|0|ON|1|ONCE none