130 Multimeter Command Reference Chapter 3
FREQuency:VOLTage:RANGe:AUTO
[SENSe:]FREQuency:VOLTage:RANGe:AUTO <mode> enables or disables the
autorange function for the signal level of frequency measurements.
Parameters
Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
If you explicitly select a range using FREQuency:VOLT:RANGe, autoranging
is turned
OFF.
Related Commands: CONFigure, PERiod:VOLT:RANGe
*RST Condition: FREQ:VOLT:RANG:AUTO ON
Example Disable Autoranging
FREQ:VOLT:RANG:AUTO OFF
Disable autorange.
FREQuency:VOLTage:RANGe:AUTO?
[SENSe:]FREQuency:VOLTage:RANGe:AUTO? returns a number to show
whether the autorange mode is enabled or disabled: “
1” = ON, “0” = OFF. The
number is sent to the output buffer.
Example Query the Autorange Mode
FREQ:VOLT:RANG:AUTO OFF
Disable autorange.
FREQ:VOLT:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
mode>
boolean OFF|0|ON|1 none