Multimeter Command Reference 157Chapter 3
:COUNt?
TRIGger:COUNt? [MIN|MAX] returns one of the following numbers to the output
buffer:
The present trigger count (1 through 50,000) if MIN or MAX are not specified.
The minimum trigger count available (1) if MIN is specified.
The maximum trigger count available (50,000) if MAX is specified.
Example Query the Trigger Count
TRIG:COUN 10
Multimeter will accept 10 triggers.
TRIG:COUN?
Query multimeter to return trigger count.
enter statement
Enter value into computer.
:DELay
TRIGger:DELay <seconds> sets the delay time between receipt of the trigger and
the start of the measurement. NOTE: This delay also occurs between each sample
when
SAMP:COUN > 1. See page 45 for a triggering process diagram.
Parameters
Comments MIN selects the minimum delay of 0 seconds for all functions. MAX selects the
maximum delay of 3600 seconds for all functions.
The trigger delay is inserted between the trigger and each measurement.
If a trigger delay is specified using the TRIG:DEL <period>,
TRIGger:DELay:AUTO is turned OFF.
The multimeter selects an automatic delay if you do not specify a trigger delay
(see the
TRIGger:DELay:AUTO command on page 158).
*RST Condition: TRIGger:DELay:AUTO ON
Example Set the Trigger Delay
TRIG:DEL .002
Wait 2ms between trigger and start of
measurement.
Parameter Name Parameter Type Range of Values Default Units
<
seconds>
numeric 0 through 3600|MIN|MAX seconds