Multimeter Command Reference 157Chapter 3
:COUNt?
TRIGger:COUNt? [MIN|MAX] returns one of the following numbers to the output
buffer:
• The present trigger count (1 through 50,000) if MIN or MAX are not specified.
• The minimum trigger count available (1) if MIN is specified.
• The maximum trigger count available (50,000) if MAX is specified.
Example Query the Trigger Count
TRIG:COUN 10
Multimeter will accept 10 triggers.
TRIG:COUN?
Query multimeter to return trigger count.
enter statement
Enter value into computer.
:DELay
TRIGger:DELay <seconds> sets the delay time between receipt of the trigger and
the start of the measurement. NOTE: This delay also occurs between each sample
when
SAMP:COUN > 1. See page 45 for a triggering process diagram.
Parameters
Comments • MIN selects the minimum delay of 0 seconds for all functions. MAX selects the
maximum delay of 3600 seconds for all functions.
• The trigger delay is inserted between the trigger and each measurement.
• If a trigger delay is specified using the TRIG:DEL <period>,
TRIGger:DELay:AUTO is turned OFF.
• The multimeter selects an automatic delay if you do not specify a trigger delay
(see the
TRIGger:DELay:AUTO command on page 158).
• *RST Condition: TRIGger:DELay:AUTO ON
Example Set the Trigger Delay
TRIG:DEL .002
Wait 2ms between trigger and start of
measurement.
Parameter Name Parameter Type Range of Values Default Units
<
seconds>
numeric 0 through 3600|MIN|MAX seconds