Multimeter Command Reference 151Chapter 3
VOLTage[:DC]:RESolution
[SENSe:]VOLTage[:DC]:RESolution <resolution> selects the resolution for DC
voltage measurements.
Parameters
Comments MINimum selects the best resolution (the smallest value) for the selected range.
MAXimum selects the worst resolution (the largest value) for the selected range.
You must select a range using VOLTage:DC:RANGe before specifying
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the range
selected during autoranging.
If autoranging is required, set the resolution using the MIN or MAX parameters
or select a specific integration time using
VOLTage:DC:NPLC.
If necessary to achieve the specified resolution, the multimeter will increase
the integration time as needed. This command overrides the results of
previously executed
VOLTage:DC:NPLC commands (the last command
executed has priority).
The VOLT:DC:RES command overrides the resolution setting from a previous
CONFigure:VOLT:DC:RES command.
Related Commands: CONFigure, VOLTage:AC:NPLC
*RST Condition: Based on the *RST values for the VOLTage:NPLC command.
Example Change the Resolution
CONF:VOLT:DC 6.25,MAX
Function: DC volts; range selected: 10V;
MAX resolution.
VOLT:DC:RANG 0.95
Range selected: 1V; MAX resolution.
VOLT:DC:RES 3E-07
Set resolution to 0.3
µ
V.
READ?
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
enter statement
Enter readings into computer.
VOLTage[:DC]:RESolution?
[SENSe:]VOLTage[:DC]:RESolution? [MIN|MAX] returns one of the following
numbers to the output buffer.
The present resolution selected if MIN or MAX is not specified. Only the
resolution values available on ranges set by the
RANGe command are
returned.
The resolution with the smallest value (i.e., the best resolution) for the selected
range if
MIN is specified.
The resolution with the largest value (i.e., the worst resolution) for the selected
range if
MAX is specified.
Parameter Name Parameter Type Range of Values Default Units
<
resolution>
numeric
resolution|MIN|MAX
volts