HP (Hewlett-Packard) E1412A Waterskis User Manual


 
150 Multimeter Command Reference Chapter 3
VOLTage[:DC]:RANGe:AUTO
[SENSe:]VOLTage[:DC]:RANGe:AUTO <mode> enables or disables the autorange
function for DC voltage measurements.
Parameters
Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
If you explicitly select a range using VOLTage:DC:RANGe, autoranging is
turned
OFF.
In order to specify an aperture time of 10
µ
s, you must select a fixed range
(e.g.,
VOLT:DC:RANG:AUTO OFF).
Related Commands: CONFigure, VOLTage:RANGe
*RST Condition: VOLT:DC:RANG:AUTO ON
Example Disable Autoranging
VOLT:DC:RANG:AUTO OFF
Disable autorange.
VOLTage[:DC]:RANGe:AUTO?
[SENSe:]VOLTage[:DC]:RANGe:AUTO? returns a number to show whether the
autorange mode is enabled or disabled: “
1” = ON, “0” = OFF. The value is sent to
the output buffer.
Example Query the Autorange Mode
VOLT:DC:RANG:AUTO OFF
Disable autorange.
VOLT:DC:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
mode>
boolean OFF|0|ON|1 none