Multimeter Command Reference 129Chapter 3
FREQuency:VOLTage:RANGe
[SENSe:]FREQuency:VOLTage:RANGe <range> selects the voltage range for the
signal level of frequency measurements.
Parameters
Comments • To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
• MIN selects the minimum range available with the VOLTage :RANGe
command: 100mV.
MAX selects the maximum range available: 300V.
• Specifying a fixed range disables the autorange mode set by the
FREQ:VOLT:RANG:AUTO command.
• The FREQ:VOLT:RANG command overrides the range setting from a previous
CONFigure:FREQuency command.
• *RST Condition: FREQ:VOLT:RANG 10
Example Set the Voltage Range for Frequency Measurements to 100V
FREQ:VOLT:RANG 100
Voltage range for frequency measurements is
100V.
FREQuency:VOLTage:RANGe?
[SENSe:]FREQuency:VOLTage:RANGe? [MIN|MAX] returns one of the
following numbers to the output buffer: 0.1, 1, 10, 100 or 300.
• The present voltage range selected if MIN or MAX is not specified.
• The minimum voltage range available (100mV) if MIN is specified.
• The maximum voltage range available (300V) if MAX is specified.
Example Query the Measurement Range
FREQ:VOLT:RANG 10
Select 10 V range.
FREQ:VOLT:RANG?
Query the present range.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
range>
numeric 100mV|1V|10V|100V|300V|
MIN|MAX
volts