HP (Hewlett-Packard) E1412A Waterskis User Manual


 
Multimeter Command Reference 129Chapter 3
FREQuency:VOLTage:RANGe
[SENSe:]FREQuency:VOLTage:RANGe <range> selects the voltage range for the
signal level of frequency measurements.
Parameters
Comments To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
MIN selects the minimum range available with the VOLTage :RANGe
command: 100mV.
MAX selects the maximum range available: 300V.
Specifying a fixed range disables the autorange mode set by the
FREQ:VOLT:RANG:AUTO command.
The FREQ:VOLT:RANG command overrides the range setting from a previous
CONFigure:FREQuency command.
*RST Condition: FREQ:VOLT:RANG 10
Example Set the Voltage Range for Frequency Measurements to 100V
FREQ:VOLT:RANG 100
Voltage range for frequency measurements is
100V.
FREQuency:VOLTage:RANGe?
[SENSe:]FREQuency:VOLTage:RANGe? [MIN|MAX] returns one of the
following numbers to the output buffer: 0.1, 1, 10, 100 or 300.
The present voltage range selected if MIN or MAX is not specified.
The minimum voltage range available (100mV) if MIN is specified.
The maximum voltage range available (300V) if MAX is specified.
Example Query the Measurement Range
FREQ:VOLT:RANG 10
Select 10 V range.
FREQ:VOLT:RANG?
Query the present range.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
range>
numeric 100mV|1V|10V|100V|300V|
MIN|MAX
volts