Multimeter Command Reference 93Chapter 3
:VOLTage:AC
CONFigure:VOLTage:AC [<range>|MIN|MAX|DEF|AUTO
[,<
resolution>|MIN|MAX|DEF]] selects the AC-coupled RMS voltage function and
allows you to specify the range and resolution.
Parameters
Comments To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range that
will accept the input.
The AUTO or DEFault option for the range parameter enables autorange. The
DEFault option for resolution defaults the integration time to 10 PLC.
The MIN and MAX parameters select the minimum or maximum values for
range:
For range:
MIN = 0.1V; MAX = 300V.
For resolution: See Table 3-4 on page 71 for valid resolution choices for
each range.
To select autorange, specify AUTO or DEF for range or do not specify a value
for the range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
To specify a MIN or MAX resolution while autoranging, you must specify
AUTO or DEFault for range; CONF:VOLT:AC DEF,MIN or
CONF:VOLT:AC DEF,MAX or CONF:VOLT:AC AUTO,MIN or
CONF:VOLT:AC AUTO,MAX (you cannot omit the range parameter). This
prevents the
MIN or MAX resolution from being interpreted as a range setting.
Example Making AC Voltage Measurements
CONF:VOLT:AC 0.54,MAX
Function: AC volts; range selected: 1A;
MAX resolution: 100
µ
A.
SAMP:COUN 3
Take 3 readings; source is IMMediate by
default.
READ?
Place multimeter in wait-for-trigger state and
make measurements; send readings to output
buffer.
enter statement
Enter readings into computer.
Parameter Name Parameter Type Range of Values Default Units
<
range>
numeric 0.1V|1V|10V|100V|300V|
MIN|MAX|DEF|AUTO
volts
<resolution>
numeric
resolution|MIN|MAX|DEF
volts