HP (Hewlett-Packard) E1412A Waterskis User Manual


 
170 Multimeter Command Reference Chapter 3
Command Description
[SENSe:] VOLTage:AC:RANGe <range>|MIN|MAX
VOLTage:AC:RANGe? [MIN|MAX]
VOLTage:AC:RANGe:AUTO OFF|ON
VOLTage:AC:RANGe:AUTO?
VOLTage:AC:RESolution <resolution>|MIN|MAX
VOLTage:AC:RESolution? [MIN|MAX]
VOLTage[:DC]:APERture .333ms|3.33ms|16.7ms|
167ms|1.67s|MIN|MAX
VOLTage[:DC]:APERture? [MIN|MAX]
VOLTage[:DC]:NPLCycles 0.02|0.2|1|10|100|
MIN|MAX
VOLTage[:DC]:NPLCycles? [MIN|MAX]
VOLTage[:DC]:RANGe <range>|MIN|MAX
VOLTage[:DC]:RANGe? [MIN|MAX]
VOLTage[:DC]:RANGe:AUTO OFF|ON
VOLTage[:DC]:RANGe:AUTO?
VOLTage[:DC]:RESolution <resolution>|MIN|MAX
VOLTage[:DC]:RESolution? [MIN|MAX]
ZERO:AUTO OFF|ONCE|ON
ZERO:AUTO?
Set range.
Query range.
Enable/disable autoranging.
Query autorange mode.
Specify resolution.
Query resolution.
Set integration time in seconds.
Query integration time (seconds).
Set integration time in PLCs.
Query integration time (PLCs).
Set range.
Query range.
Enable/disable autoranging.
Query autorange mode.
Specify resolution.
Query resolution.
Enable/disable autozero mode.
Query autozero mode.
STATus :PRESet
:QUEStionable:CONDition?
:QUEStionable:ENABle <unmask>
:QUEStionable:ENABle?
:QUEStionable[:EVENt]?
Sets all bits of enable register to “0”.
Query the questionable condition register.
Sets conditions in enable register.
Query the questionable enable register.
Query the questionable event register.
SYSTem :ERRor?
:VERSion?
Return error number/message from error queue.
Return the multimeter's SCPI version.
TRIGger :COUNt <number>|MIN|MAX|INFinite
:COUNt? [MIN|MAX]
:DELay <seconds>|MIN|MAX
:DELay? [MIN|MAX]
:DELay:AUTO OFF|ON
:DELay:AUTO?
:SOURce BUS|IMMediate|EXTernal|TTLTrg0-7
:SOURce?
Set number of triggers or scans.
Query trigger count.
Set delay between trigger and start of
measurement.
Query trigger delay.
Enable/disable automatic trigger delay.
Query automatic trigger delay mode.
Specify trigger source.
Query trigger source.