Multimeter Command Reference 137Chapter 3
PERiod:VOLTage:RANGe
[SENSe:]PERiod:VOLTage:RANGe <range> selects the voltage range for the
signal level of period measurements.
Parameters
Comments To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
MIN selects the minimum range available with the PERiod:VOLTage:RANGe
command: 100mV.
MAX selects the maximum range available: 300V.
Specifying a fixed range disables the autorange mode set by the
PER:VOLT:RANG:AUTO command.
*RST Condition: PER:VOLT:RANG 10
Example Set the Voltage Range for Period Measurements to 100V
PER:VOLT:RANG 100
Voltage range is 100 V.
PERiod:VOLTage:RANGe?
[SENSe:]PERiod:VOLTage:RANGe? [MIN|MAX] returns one of the following
numbers to the output buffer:
0.1, 1, 10, 100 or 300 corresponding to the range set.
MIN returns 0.1.
MAX returns 300.
Example Query the Period Voltage Range
PER:VOLT:RANG?
Query the voltage range for period
measurements.
enter statement
Enter response into computer.
Parameter Name Parameter Type Range of Values Default Units
<
range>
numeric 100mV|1V|10V|100V|300V|
MIN|MAX
volts