HP (Hewlett-Packard) E1412A Waterskis User Manual


 
122 Multimeter Command Reference Chapter 3
CURRent[:DC]:APERture
[SENSe:]CURRent[:DC]:APERture <number> sets the integration time in seconds
for dc current measurements. Values are rounded up to the nearest aperture time
shown in the following table.
Parameters
Comments MIN sets the aperture time to 0.333 ms. MAX sets the aperture time to
1.66667 seconds (60Hz) or 2 seconds (50Hz).
Setting the aperture time also sets the integration time in power line cycles
(PLCs) and the resolution. For example, an aperture time of 16.7ms (60Hz line
frequency) sets an integration time of 1 PLC. The corresponding resolution
depends on the function and range you select.
The CURR:DC:APER command overrides the results of previously executed
CURR:DC:NPLC and CURR:DC:RES commands. The last command executed
has priority.
The greater the aperture time, the greater the normal mode rejection (and the
lower the reading rate).
Related Commands: CALibration:LFRrequency
*RST Condition: CURR:DC:APER 0.166667 seconds (60Hz) or
CURR:DC:APER 0.20000 (50Hz)
Example Set an Aperture Time of 16.7ms
CURR:APER 16.7E-03
Aperture time is 16.7 ms.
CURRent[:DC]:APERture?
[SENSe:]CURRent[:DC]:APERture? [MIN|MAX] returns one of the following
numbers to the output buffer.
The present aperture time in seconds if MIN or MAX is not specified.
The minimum aperture time available (.333 ms) if MIN is specified.
The maximum aperture time available (1.67 s @ 60Hz; 2 s @ 50Hz) if MAX is
specified.
Example Query the Aperture Time
CURR:APER 167E-03
Aperture time is 167ms.
CURR:APER?
Query multimeter to return aperture time.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
number>
numeric 0.333ms|3.33ms|16.7ms|
167ms|1.67s|MIN|MAX
seconds