HP (Hewlett-Packard) E1412A Waterskis User Manual


 
116 Multimeter Command Reference Chapter 3
[SENSe:]
The [SENSe:] command subsystem is most commonly used with CONFigure to
change specific “low-level” measurement parameters.
[SENSe:] enables you to
change the following measurement parameters, predefined by the
CONFigure
command, without completely reconfiguring the multimeter.
-- Function, Range and Resolution
-- Aperture Time and Number of Power Line Cycles
-- Bandwidth
-- Autozero
Subsystem Syntax [SENSe:]
FUNCtion “CURRent:AC”
FUNCtion “CURRent[:DC]”
FUNCtion “FREQuency”
FUNCtion “FRESistance”
FUNCtion “PERiod”
FUNCtion “RESistance”
FUNCtion “VOLTage:AC”
FUNCtion “VOLTage[:DC]”
FUNCtion “VOLTage[:DC]:RATio”
FUNCtion?
CURRent
:AC:RANGe <
range>|MIN|MAX
:AC:RANGe? [MIN|MAX]
:AC:RANGe:AUTO OFF|ON
:AC:RANGe:AUTO?
:AC:RESolution <
resolution> |MIN|MAX
:AC:RESolution? [MIN|MAX]
[:DC]:APERture .333ms|3.33ms|16.7ms|167ms|1.67s|MIN|MAX
[:DC]:APERture? [MIN|MAX]
[:DC]:NPLCycles .02|.2|1|10|100|MIN|MAX
[:DC]:NPLCycles? [MIN|MAX]
[:DC]:RANGe <
range>|MIN|MAX
[:DC]:RANGe? [MIN|MAX]
[:DC]:RANGe:AUTO OFF|ON
[:DC]:RANGe:AUTO?
[:DC]:RESolution <
resolution>|MIN|MAX
[:DC]:RESolution? [MIN|MAX]
DETector
:BANDwidth 3|20|200|MIN|MAX
:BANDwidth? [MIN|MAX]
FREQuency
:APERture 0.01|0.1|1|MIN|MAX
:APERture? [MIN|MAX]
:VOLTage:RANGe <
range>|MIN|MAX
:VOLTage:RANGe? [MIN|MAX]
:VOLTage:RANGe:AUTO OFF|ON
:VOLTage:RANGe:AUTO?