138 Multimeter Command Reference Chapter 3
PERiod:VOLTage:RANGe:AUTO
[SENSe:]PERiod:VOLTage:RANGe:AUTO <mode> enables or disables the
autorange function for the signal level of period measurements.
Parameters
Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
If you explicitly select a range using PERiod:VOLT:RANGe, autoranging is
turned
OFF.
Related Commands: CONFigure, FREQuency:VOLT:RANGe
*RST Condition: PER:VOLT:RANG:AUTO ON
Example Disable Autoranging
PER:VOLT:RANG:AUTO OFF
Disable autorange.
PERiod:VOLTage:RANGe:AUTO?
[SENSe:]PERiod:VOLTage:RANGe:AUTO? returns a number to show whether
the autorange mode is enabled or disabled: “
1” = ON, “0” = OFF. The number is sent
to the output buffer.
Example Query the Autorange Mode
PER:VOLT:RANG:AUTO OFF
Disable autorange.
PER:VOLT:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
mode>
boolean OFF|0|ON|1 none