HP (Hewlett-Packard) E1412A Waterskis User Manual


 
74 Multimeter Command Reference Chapter 3
:AVERage:AVERage?
CALCulate:AVERage:AVERage? reads the average of all readings taken since
AVERage was enabled (CALC:FUNC AVER and CALC:STAT ON commands). The
average value is cleared when
AVERage is enabled, when power is removed, or after
the multimeter is reset. The average value is stored in volatile memory.
Example Query the Average of All Readings Taken Since the AVERage Math Operation
was Enabled
CALC:AVER:AVER?
Query the average of all readings.
:AVERage:COUNt?
CALCulate:AVERage:COUNt? reads the number of readings taken since
AVERage was enabled (CALC:FUNC AVER and CALC:STAT ON commands). The
count value is cleared when
AVERage is enabled by the CALC:FUNC AVER and
CALC:STAT ON commands, when power has been off, or after a remote interface
reset. The number of readings taken is stored in volatile memory.
Example Query the Number of Readings Since the AVERage Math Operation was
Enabled
CALC:COUN?
Query number of readings.
:AVERage:MAXimum?
CALCulate:AVERage:MAXimum? reads the maximum value found from an
AVERage operation. The max value is cleared when AVERage is enabled
(
CALC:FUNC AVER and CALC:STAT ON commands), when power is removed, or
after the multimeter is reset. The maximum value is stored in volatile memory.
Example Query the Maximum Value Found During an AVERage Math Operation
CALC:AVER:MAX?
Query the max value.
:AVERage:MINimum?
CALCulate:AVERage:MINimum? reads the minimum value found from an
AVERage function operation. The min value is cleared when AVERage is enabled
(
CALC:FUNC AVER and CALC:STAT ON commands), when power is removed, or
after the multimeter is reset. The minimum value is stored in volatile memory.
Example Query the Minimum Value Found During an AVERage Math Operation
CALC:AVER:MIN?
Query the min value.