Multimeter Command Reference 159Chapter 3
:DELay:AUTO?
TRIGger:DELay:AUTO? returns a number to show whether the automatic trigger
delay mode is on or off: “
1” = ON, “0” = OFF. The number is sent to the output
buffer.
Example Query the Trigger Delay Mode
TRIG:DEL:AUTO OFF
Disable automatic trigger delay.
TRIG:DEL:AUTO?
Query multimeter to return trigger delay mode.
enter statement
Enter value into computer.
Default Trigger Delays for DC Voltage and DC Current (all ranges):
Integration Time Trigger Delay
NPLC ≥11.5ms
NPLC <1 1.0ms
Default Trigger Delays for 2-Wire and 4-Wire Resistance:
Range Trigger Delay
(For NPLC ≥1)
Trigger Delay
(For NPLC <1)
100Ω 1.5ms 1.0ms
1kΩ 1.5ms 1.0ms
10kΩ 1.5ms 1.0ms
100kΩ 1.5ms 1.0ms
1MΩ 1.5ms 10ms
10MΩ 100ms 100ms
100MΩ 100ms 100ms
Default Trigger Delays for AC Voltage and AC Current (all ranges):
AC Filter Trigger Delay
3Hz - 300kHz filter (Slow) 7.0sec
20Hz - 300kHz filter (Medium) 1.0sec
200Hz - 300kHz filter (Fast) 600ms
Default Trigger Delay for Frequency and Period:
1.0s